Magnet Wire Testers
The test cage is equipped with a plastic door so that the operation of the test may be observed. This door is fitted with a safety interlock switch so that the operator cannot receive an electrical shock. Mercury* cups are machined in a nylon bar, which plugs into sockets within the cage. This permits easy removal of the cups for filling, emptying and cleaning.
The maximum output voltage is 20kv, and there is an adjustable (automatic) rate- of-rise of voltage to the specimen. Testing always begins at zero volts. When breakdown occurs, the reading of the voltage at breakdown remains on the voltmeter, although the high tension is automatically disconnected immediately.
THIS METER MEETS THE FOLLOWING STANDARDS:
- ASTM-D149, D1676, D3353
ANSI-C68.1
IEEE-Publication No, 4
|
SPECIFICATIONS |
|
120V, 60 Hz, single phase, 10 A |
|
0-20 kV ac adjustable rate of rise 100-1000 V/s |
|
adjustable by a front panel
control from about 2.5 to 60mA |
|
memory digital voltmeter reads to 19.99kV |
|
provision for external safety
interlock switches provision for external high-voltage-on lamp provision for external fault lamp |
|
50W x 60H X45D cm (approx.) (20W x 24H x 18 in) |
|
40 kg (85 lb) (approx.) |
For production use we suggest our AC Sparker, Model AVS-300 |
N/A, 30 Progress Avenue, Scarborough, ON M1P 4W8. Tel: 416-299-6666; Fax: 416-299-8398.